14-08 Imagen cardíaca por RM

One of the biggest challenges for MR imaging is the heart [⇒ Bradley 1992]. It not only contracts, it also moves due to respiration; it contains flowing blood, and its axes are not orthogonal to the rest of the body (Figure 14-23 and Figure 14-24).

The latter point is not a restriction in MR imaging because it is basically a three-dimensional technique; the former points limited cardiac MR imaging strongly, because it used to be a slow modality and the time resolution required for cardiac imaging is less than 50 ms.


Figura 14-23:
Anatomic sketch of the human heart with depiction of the long axis (blue) and the short axis (green).

Figura 14-24:
Morfología cardíaca: Corte tranverso a través del corazón a nivel de válvula aórtica (imágenes con sangre negra).


Thus, cardiac imaging requires synchronization of data acquisition and the different kinds of motion of the heart, otherwise images are degraded. Three types of synchronization are possible: gating, triggering, and slice following.

Gating means opening a gate or a time window during which the data ac­qui­si­tion can run freely. The acquisition is stopped when the gate is closed, and con­ti­nued as the gate opens again. The time window during which the gate is open is not necessarily the same from gate to gate. Opening and closing of the gate is con­trol­led by a physiological monitor, such as a chest elevation monitor from respiratory gating.

Triggering means receiving a signal that starts one or several pulses. The num­ber of pulses is the same every time. For good timing, the trigger pulse should stem from an easily definable and exact physiological incident, like the peak of the R-wave in the QRS-complex of an electrocardiogram.

Distortions of the detected QRS-complex are readily introduced by gradient pulsing or by material in ECG electrodes. Some problems cannot usually be avoided, such as the elevation of the T-wave due to the flow of blood through a magnetic field (see Chapter 18). This effect increases with field strength and with dobutamine stimulation of the heart in stress examinations. Since the ele­va­tion of the T-wave can give falsely detected R-waves, the triggering parameter should be set well ahead of the QRS-complex. There are some advanced ECG sys­tems that use an out-of-magnet recorded vector ECG as a mask to get an un­dis­tur­bed ECG inside the magnet during imaging.

Slice following is a technique where both the movement of the diaphragm du­ring respiration and the heart during contraction is monitored by the system. Through checking the positions of anatomical landmarks immediately before and after the acquisition of an imaging profile, the displacement of the acquired profile can be checked (and discarded if outside given borders) and the dis­place­ment of the following profile can be predicted. Slice following is often used in com­bi­na­tion with triggering to achieve better temporal resolution.

Measurement of myocardial wall motion becomes possible by myocardial tagging. Tagging entails labeling a strip in the myocardium by magnetic sa­tu­ra­tion. Thus wall motion can be directly monitored (Figure 14-25) [⇒ Zerhouni].


Figure 14-25:
Heart examination with spin tagging. Myocardial wall motion can be followed because the spins of the dark mesh were selectively excited. They re­tain a state of excitation different from their neigh­bors for the duration of T1.


14-08-01 Static Studies

For the depiction of cardiac and great vessel morphology, static studies in se­ve­ral phases of the heart cycle are performed, usually as SE or RSE studies; pri­ma­ri­ly as black blood images. They should be performed as a multi-slice, multi-phase, double-oblique angulated acquisition, where special attention must be paid to the patient’s heart frequency. Care should be taken about the chosen imag­ing planes; four-chamber views are generally coronal, but long-axis views can be either sagittal or transverse, whereas short-axis views can be either trans­ver­se or sagittal.


14-08-02 Flow Studies

Bulk flow either from shunts across the septum, regurgitant jets through valves closing insufficiently, or just through lumina and vessels can be visualized by gradient-echo techniques.

These studies must have reasonable temporal resolution to describe the dif­fe­rent phases of the heart cycle, typically 16 or more, depending on the clinical question (Figure 14-26). The results will yield an image with muscular tissue in gray, static liquid in white and high-velocity jets in black (signal void). Con­se­quent­ly, both ordinary flow and regurgitant jets are seen, but cannot be quan­ti­fied immediately. One quantification method being employed is to measure the area (or volume) of the regurgitant jet (signal void) and compare it to the area (or volume) of the chamber.


Figure 14-26:
Patient with several transient ischemic at­tacks. Gradient-echo images four mil­li­me­ters above the aortic valve during different phases of the cardiac cycle show a pen­du­lat­ing thrombus.


Another technique tracks the signal intensity of the blood in the chamber du­ring the cardiac cycle. The total signal intensity increases in normal patients du­ring systole, but decreases markedly in patients with regurgitation. The per­cen­tage of decrease is found to be dependent on the severity of the regurgitation. An­gio­gra­phic techniques such as flow quantification can also be utilized and then net flow through an orifice can be accurately quantified.


14-08-03 Clinical Applications

To date, the combination of SE/RSE and fast gradient-echo imaging of the heart has been found to be the most efficient method to image the heart by magnetic resonance. The goal of such an examination is to combine the evaluation of mor­pho­lo­gy with functional features. In clinical routine, imaging time should not ex­ceed thirty minutes; image processing, and particularly interpretation, will take longer. Great care has to be put into planning and optimizing a heart exa­mi­na­tion, and certain trade-offs should be realized. A good and detailed review was published by Dudley J. Pennell in 2010 [⇒ Pennell]; helpful clinial guidelines are given by a collaborating team of several cardiac imaging societies [⇒ Kramer et al. 2013].

In spin-echo images, regional abnormal wall motion or abnormal wall thick­ness is seen, as well as pericardial disturbances. Fatty deposits in the myo­car­dium of the right ventricle and intra- or extracardiac tumors, together with crypts and ducts, are also generally found with ease.

Gradient-echo images tend to give somewhat poorer edge description of the endocardium, but a good overview of different hypertrophies. Specially designed gradient-echo sequences like True FISP and Balanced FFE do, however, give ex­cel­lent blood-tissue contrast. Furthermore, both restricted and dilated cavities, insufficient valves and tracts are easily seen. The general flow pattern and the to­tal overview of the heart add to the general understanding of cardiac per­for­man­ce.


14-08-04 Advanced Techniques

Both first-pass contrast uptake and late-enhancement imaging are gaining ground in cardiac diagnostic imaging. These kinds of contrast studies are used to evaluate cardiac perfusion and perfusion reserve and to qualify the possible via­bi­li­ty of cardiac tissue. Coronary artery imaging is also progressing using mul­ti­ple different 2D- and 3D-imaging sequences (RSE, rapid GRE, SE-EPI, rapid GRE-EPI).

The resolution reached has been good enough to persistently describe 7-10 cm of the main coronary arteries (RAD, LAD, and circumflex), but not the col­la­te­rals (Figure 14-27). This resolution is sufficient to evaluate the patency of grafts and MR coronary artery imaging can be used in patients with severe ana­phy­lac­tic reactions to contrast media, but it is still not a sufficiently robust screening tool.

Figure 14-27:
Depiction of the coronary arteries without contrast agent application by spiral 3D ra­pid gradient-echo using navigator tech­ni­ques and diastolic gating.


Cardiac MR imaging is considered the method-of-choice for examinations of congenital heart disease, acquired diseases of the great vessels and tumors in­fil­trat­ing or close to the heart and valuable in a number of other clinical areas such as cardiomyopathies, pericardial diseases and post-transplantation exa­mi­na­tions.

spaceholder 600 spaceholder 600

LogoTop
LogoBottom
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2